Gastec 192 Quick-measuring Detector tubes Acrylonitrile CH2:CHCN (0.2~32ppm)
Manufacturer: Gastec Model: 192 - Request a Quotation
- Contact
See more: Equipment Calibration-Inspection & Repair Service
Call for the best price
Hanoi city: (024) 35.381.269
Danang city: (023) 63.747.711
Bac Ninh city: (0222) 730.39.68
Hai Phong city: (0225) 730.03.89
HCM city: (028) 38.119.636
Dong Nai: 0345.689.536
Indicator: 0.5 to 10 ppm
Measuring Range:
+ 0.2 to 0.5 ppm , Number of Pump Strokes 4
+ 0.5 to 10 ppm, Number of Pump Strokes 2
+ 10 to 32 ppm , Number of Pump Strokes 1
Sampling Time: 2 minute per pump stroke
Colour Change: Yellow → Red
Reaction Principle:
+ CH2:C(CH3)CN+ Cr6+ + H2SO4 → HCN
+ 2HCN + HgCl2 → 2HCl+Hg(CN) 2
+ HCl+Base → Chloride
Coefficient of Variation: 10% (for 0.5 to 2 ppm), 5% (for 2 to 10 ppm)
Nr. of tests: 5
Pump: GV-100S/110S
Measuring Range |
|
|||
Number of Pump Strokes |
|
|||
Correction Factor |
|
|||
Sampling Time | 2 minutes per pump stroke | |||
Detecting Limit | 0.1 ppm (n=4) | |||
Colour Change | Yellow → Red | |||
Reaction Principle | CH2:C(CH3)CN+ Cr6+ + H2SO4 → HCN 2HCN + HgCl2 → 2HCl+Hg(CN) 2 HCl+Base → Chloride |
|||
Coefficient of Variation | 10% (for 0.5 to 2 ppm), 5% (for 2 to 10 ppm) | |||
Shelf Life | 3 Years | |||
Corrections for temperature & humidity | Unnecessary | |||
Store the tubes at cool and dark place. |
Substance | Concentration | Interference | Change colour by itself to |
Nitriles (C3) | + | Red | |
Acetone cyanohidrin | + | Red | |
Alcohols, Esters, Ketones | 20 ppm | No | |
Aromatic hydrocarbons | 20 ppm | No | |
Hydrogen chloride, Hydrogen cyanide |
No | No |
Calibration gas generation | Diffusion tube method |
TLV-TWA | TLV-STEL | Explosive range |
1 ppm(2014) | - | - |