Gastec 4LK Quick-measuring Detector tubes Hydrogen sulphide H2S
Manufacturer: Gastec Model: 4LK Origin: Japan - Contact
- In stock: 1 Unit
See more: Equipment Calibration-Inspection & Repair Service
Call for the best price
Hanoi city: (024) 35.381.269
Danang city: (023) 63.747.711
Bac Ninh city: (0222) 730.39.68
Hai Phong city: (0225) 730.03.89
HCM city: (028) 38.119.636
Dong Nai: 0932.160.940
Indicator: 2 to 20 ppm
Measuring Range:
+ 1 to 2 ppm, Number of Pump Strokes 2
+ 2 to 20 ppm, Number of Pump Strokes 1
+ 20 to 40 ppm, Number of Pump Strokes 1/2
Sampling Time: 1 minute per pump stroke
Colour Change: White → Brown
Reaction Principle: H2S + Pb(CH3COO)2 → PbS + 2CH3COOH
Coefficient of Variation: 10% (for 2 to 6 ppm), 5% (for 6 to 20 ppm)
Nr. of tests: 10
| Measuring Range |
|
|||
| Number of Pump Strokes |
|
|||
| Correction Factor |
|
|||
| Sampling Time | 1 minute per pump stroke | |||
| Detecting Limit | 0.25 ppm (n=2) | |||
| Colour Change | White → Brown | |||
| Reaction Principle | H2S + Pb(CH3COO)2 → PbS + 2CH3COOH | |||
| Coefficient of Variation | 10% (for 2 to 6 ppm), 5% (for 6 to 20 ppm) | |||
| Shelf Life | 3 Years | |||
| Corrections for temperature & humidity | Unnecessary | |||
| Store the tubes at cool and dark place. | ||||
| Substance | Concentration | Interference | Change colour by itself to |
| Mercaptans | No | No | |
| sulphur dioxide | 1/2 | + | No |
| Nitrogen dioxide | 1/4 | No | |
| Ozone | 10 ppm | Bleaching | No |
| Calibration gas generation | Permeation tube method |
| TLV-TWA | TLV-STEL | Explosive range |
| 1 ppm | 5 ppm | 4.0 to 44% |
- Quality Engagement
- Easy change and return
- Delivery Avaliable
- Favorable payment