HIOKI CT6845 AC/DC CURRENT PROBE
Manufacturer: HIOKI Model: CT6845 - Request a Quotation
- Contact
See more: Equipment Calibration-Inspection & Repair Service
Call for the best price
Hanoi city: (024) 35.381.269
Danang city: (023) 63.747.711
Bac Ninh city: (0222) 730.39.68
Hai Phong city: (0225) 730.03.89
HCM city: (028) 38.119.636
Dong Nai: 0345.689.536
Rated current: 500 A AC/DC
Max. allowable input:
- 1000A continuous (requires derating at frequency or temperature)
- DC to 100 kHz
- Phase: DC to 100 kHz
Amplitude and phase accuracy:
- DC ±0.3 % rdg. ±0.02 % f.s. (Phase: Not defined)
- DC < f ≤ 100 Hz ±0.3 % rdg. ±0.01 % f.s. (Phase: ±0.1 deg)
- Defined to 100 kHz
Power consumption : 7 VA max. (at 500 A/55 Hz, ±12 V power requirement)
Output voltage: 4 mV/A (voltage output with the Sensor Unit 9555-10, use with a device having a 1 MΩ input resistance or higher)
Core diameter: φ 50 mm (1.79 in)
Power supply: DC ±11 V to ±15 V (Power suppled via the 9555-10, which supports 100 to 240 V AC)
Dimensions and mass: 238 mm (9.37 in)W × 116 mm (4.57 in)H × 35 mm (1.38 in)D, cord length: 3 m (9.84 ft), 870 g (30.7 oz)
Accessories: Instruction manual ×1, Mark bands ×6, Carrying Case
■ Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
Rated current | 500 A AC/DC |
---|---|
Max. allowable input | 1000A continuous (requires derating at frequency or temperature) |
Frequency characteristics | DC to 100 kHz Phase: DC to 100 kHz |
Amplitude and phase accuracy | DC ±0.3 % rdg. ±0.02 % f.s. (Phase: Not defined) DC < f ≤ 100 Hz ±0.3 % rdg. ±0.01 % f.s. (Phase: ±0.1 deg) Defined to 100 kHz |
Power consumption | 7 VA max. (at 500 A/55 Hz, ±12 V power requirement) |
Output voltage | 4 mV/A (voltage output with the Sensor Unit 9555-10, use with a device having a 1 MΩ input resistance or higher) |
Core diameter | φ 50 mm (1.79 in) |
Operating temperature, humidity | -40 °C to +85 °C (-40 °F to 185 °F), 80% rh or less (with no condensation) |
Power supply | DC ±11 V to ±15 V (Power suppled via the 9555-10, which supports 100 to 240 V AC) |
Dimensions and mass | 238 mm (9.37 in)W × 116 mm (4.57 in)H × 35 mm (1.38 in)D, cord length: 3 m (9.84 ft), 870 g (30.7 oz) |
Accessories | Instruction manual ×1, Mark bands ×6, Carrying Case |