IETLAB HPRS-HIPOT-LOAD (Fluke 5320A-Load Hipot Replacement) Hipot Current Calibration Load Resistors (±10%; 10 kΩ tới 5 MΩ )
Manufacturer: IET LAB Model: HPRS-HIPOT-LOAD Guarantee: 12 Month - Request a Quotation
- Contact
See more: Equipment Calibration-Inspection & Repair Service
Call for the best price
Hanoi city: (024) 35.381.269
Danang city: (023) 63.747.711
Bac Ninh city: (0222) 730.39.68
Hai Phong city: (0225) 730.03.89
HCM city: (028) 38.119.636
Dong Nai: 0345.689.536
(Full specifications can be found on datasheet)
8 Resistances from 10 kΩ to 5 MΩ standard
Standard accuracy: ±10% at 23°C
Power rating : up to 300 W at 100 kΩ
Voltage rating: 1,400 to 5,500 V depending upon resistance value
Continuous operation at rated power
Kel-F washers on all terminals for low leakage current and high dielectric strength
Custom Values Available
BASIC SPECIFICATIONS
(See datasheet for full specifications)
Resistance Value |
Maximum Voltage |
Max power |
Resistor Type |
10 kΩ |
1400 V |
196 W |
Tubular ceramic |
35 kΩ |
2000 V |
110 W |
|
50 kΩ |
2000 V |
80 W |
|
100 kΩ |
5500 V |
300 W |
|
250 kΩ |
5500 V |
120 W |
|
500 kΩ |
5500 V |
60 W |
|
1 MΩ |
5500 V |
30 W |
|
5 MΩ |
5500 V |
5 W |
Thick film |
Adjustment to Nominal: ±10% at 23°C and 95% confidence level; traceable to SI
Terminals: 9 five-way binding posts with Kel-F washers and one ground post electrically connected to case
Dimensions: 42.7 cm W x 12.4 cm H x 31.5 cm D (16.8" x 5.2" x 12.4")
Weight: 4.5 kg (10 lb), nominal
The IET HPRS-Hipot-Load is an excellent replacement for the Fluke 5320A-Load Hipot current calibration load resistors. The IET high power resistive load can be used with the Fluke 5320A Multifunction Electrical Tester Calibrator or standalone. It provides a high power and high voltage load to various hipot tester to create the leakage current needed to calibrate various manufactures hipot testers.
The HPRS-HIPOT-LOAD consist of a series of 8 resistances from 10 kΩ to 5 MΩ. The resistances are in series between the Lo and the 5 MΩ terminals with taps connected to the other resistance values.
The 8 resistance values were chosen to mimic the Fluke 5320A-load. IET can build the HPRS Hipot Load with a range of resistance values, accuracies, power ratings and maximum voltage depending upon your requirement.
Typical Application: The hipot tester is connected to the various calibration load resistors during the calibration process. The voltage across the load and the current flowing through the load can then be measured. The measured voltage and current can then be compared with the displayed value from the hipot tester during calibration or as part of the adjustment procedure.