NGI N6212-60-100 Single-Channel Medium Power DC Electronic Load (60V, 100A, 1200W)

Voltage: 60V

Current: 100A

Power: 1200W

Min operating voltage: 2V@100A

CC mode

Range: 0~10A/ 0~100A

Setting Resolution: 1mA/ 10mA

Setting Accuracy (23±5℃): Low range:0.1%+0.1%F.S.,/ High range:0.1%+0.15%F.S

CV mode

Range: 0~6V/ 0~60V

Setting Resolution: 0.1mV/ 1mV

Setting Accuracy: (23±5℃) 0.05%+0.1%F.S.

CP mode

Range: 0~1200W

Setting Resolution: 10mW

Setting Accuracy (23±5℃): 0.5%+1%F.S

CR mode

Range: 0.02~60Ω/ 0.3~600Ω

Setting Resolution: 16 bits

Setting Accuracy (23±5℃): 0.35%+52mS/ 0.35%+5.2mS

Slew Rate

Current: 1.6~100A/ms/ 100~5000A/ms

Voltage: 0.5~25V/ms/ 25~250V/ms

Power: 1.6~100A/ms/ 100~5000A/ms

Resistance: 1.6~100A/ms/ 100~5000A/ms

Voltage Measurement

Range: 0~6V/ 0~60V

Readback Accuracy (23±5℃): 0.05%+0.05%F.S.

Current Measurement

Range: 0~10A / 0~100A

Readback Accuracy (23±5℃):0.05%+0.1%F.S.

Power Measurement

Range: 0~1200W

Readback Accuracy (23±5℃): 0.5%+1%F.S. 

Interface: LAN/RS232

AC Input: Single phase, 220V AC±10%, frequency 47Hz~63Hz

Sampling Frequency: 25Hz

Communication Response Time: ≤10ms

Operating Environment: Altitude: <2000m, relative humidity: 5%~90%RH(non-condensing), atmospheric pressure: 80~110kPa

Temperature: Operating temperature: 0℃~40℃, storage temperature: -20℃~60℃

Net weight: Approx. 13kg

Dimension: 2U, 88.0(H)*482.0(W)with handle*507.0(D)mm